头颈部肿瘤调强放疗中摆位偏差的测量与分析
http://www.100md.com 《第四军医大学学报》 2006年第12期
头颈部肿瘤;摆位误差;电子射野影像装置;数字重建图像,,头颈部肿瘤;摆位误差;电子射野影像装置;数字重建图像,0引言,1材料和方法,2结果,3讨论,【参
     【Abstract】 AIM: To measure the setup error in the intensity modulated radiation therapy (IMRT) for headneck neoplasms, and to analyze the margin between the clinical target volume (CTV) and planning target volume (PTV). METHODS: Seventysix patients with headneck neoplasms were choosed at random. The portal films were screened using the electronic portal imaging device (EPID), and the error was analyzed through comparing the portal film with the digitally reconstruced radiographs (DRR) in the treatment planning system (TPS). RESULTS: The setup errors were (-0.62±1.46) mm,(-0.41±1.54) mm,(-0.31±1.67) mm and the margins were 2.27 mm, 1.87 mm, 1.98 mm in lateral, cranialcaudal and ventraldorsal directions respectively. CONCLUSION: For the patients with headneck neoplasms undergoing the IMRT, the margins between the CTV and the PTV should be 2.5 mm in the lateral direction, and 2 mm in cranialcaudal and ventraldorsal directions. Moreover, with the treatment time goes by, the error varied irregularly. ......