结扎速血管闭合系统闭合实验狗血管的可靠性
http://www.100md.com
王永来, 陈 双, 苏 华, 刘彩刚
血管;止血;结扎速血管闭合系统,王永来,陈双,苏华,刘彩刚,通讯作者:,电话:,收稿日期:,接受日期:,Reliabilityofhemostasiswithligasurevesselsealingsysteminexperimentaldogs
参见附件(1kb)。
王永来, 陈双, 苏华, 刘彩刚, 中国医科大学附属盛京医院妇产科 辽宁省沈阳市 110004
通讯作者: 王永来, 110004, 辽宁省沈阳市和平区三好街36号, 中国医科大学附属盛京医院妇产科. yonglai_wang@sina.com
电话: 024-25943605
收稿日期: 2006-12-12 接受日期: 2007-01-10
Reliability of hemostasis with ligasure vessel sealing system in experimental dogs
Yong-Lai Wang, Shuang Chen, Hua Su, Cai-Gang Liu
Yong-Lai Wang, Shuang Chen, Hua Su, Cai-Gang Liu,Department of Obstetrics and Gynecology, Shengjing Hospital of China Medical University, Shenyang 110004, Liaoning Province, China
Correspondence to:Yong-Lai Wang, Department of Obstetrics and Gynecology, Shengjing Hospital of China Medical University, 36 Sanhao Street, Heping District, Shenyang 110004, Liaoning Province,China. yonglai_wang@sina.com
Received:2006-12-12 Accepted:2007-01-10
Abstract
AIM:To study the reliability and tissue damage of ligasure vessel sealing system during vascular occlusion.
METHODS: Ligasure vessel sealing system was used to seal and cut off the dog blood vessel of different locations and diameters. The diameter of the vessel was measured by a caliper before and after sealing and cutting off, respectively. The bursting pressures were recorded after sealed, and the vascular pathology was performed after the operation.
RESULTS: A total of 12 vessels were sealed successfully by ligasure vessel sealing system, and the success rate was 100%. The mean diameters of the sealed artery and vein were 3.26 mm (1.75-5.22 mm) and 5.42 mm (2.40-9.08 mm), while the bursting pressures were 472 mmHg (380-560 mmHg) and 378 mmHg (250-490 mmHg), respectively. The pathological changes of the sealed vessels caused by ligasure vessel sealing system mainly presented as acute thermal and pressure-induced injury.
CONCLUSION: Ligasure vessel sealing system can effectively seal the vessels, especially the ones 3 mm or larger in diameter.
Key Words: Blood vessel; Hemostasis; Ligasure vessel sealing system
Wang YL, Chen S, Su H, Liu CG. Reliability of hemostasis with ligasure vessel sealing system in experimental dogs. Shijie Huaren Xiaohua Zazhi 2007;15(8):890-892
摘要
目的: 研究结扎速(ligasure)血管闭合系统闭合实验狗血管的可靠性及组织损伤程度.
方法: 应用结扎速血管闭合系统切断和闭合实验狗体内不同部位与直径的血管, 切断和闭合血管前用游标尺测量血管直径, 闭合后测定血管破裂时的压力, 术后血管行组织病理学检查.
结果: 结扎速血管闭合系统闭合12条动、静脉, 闭合成功率均为100%, 平均动静脉直径分别是3.26 mm(1.75-5.22 mm)和5.42 mm(2.40-9.08 mm). 成功闭合血管的平均动、静脉破裂压分别是472 mmHg(380-560 mmHg), 378 mmHg(250-490 mmHg). 术后损伤的主要病理改变为组织的急性热和压力损伤.
结论: 结扎速血管闭合系统闭合直径3 mm或更粗的血管有较高的可靠性.
关键词: 血管; 止血; 结扎速血管闭合系统
王永来, 陈双, 苏华, 刘彩刚. 结扎速血管闭合系统闭合实验狗血管的可靠性. 世界华人消化杂志 2007;15(8):890-892
0 引言
随着腹腔镜技术的飞速发展, 传统的切开及止血方法已经满足不了腹腔镜手术的要求. 目前有关各种止血设备闭合血管的可靠性及组织损伤程度的实验研究少见报道. 结扎速血管闭合系统(ligasure vessel sealing system)是一种新型止血设备, 其独特之处是以组织反应发生器作为电流和电压的能量来源. 我们通过研究结扎速血管闭合系统闭合血管的可靠性及组织损伤程度, 为进行腹腔镜手术的临床医生提供科学依据 ......
王永来, 陈双, 苏华, 刘彩刚, 中国医科大学附属盛京医院妇产科 辽宁省沈阳市 110004
通讯作者: 王永来, 110004, 辽宁省沈阳市和平区三好街36号, 中国医科大学附属盛京医院妇产科. yonglai_wang@sina.com
电话: 024-25943605
收稿日期: 2006-12-12 接受日期: 2007-01-10
Reliability of hemostasis with ligasure vessel sealing system in experimental dogs
Yong-Lai Wang, Shuang Chen, Hua Su, Cai-Gang Liu
Yong-Lai Wang, Shuang Chen, Hua Su, Cai-Gang Liu,Department of Obstetrics and Gynecology, Shengjing Hospital of China Medical University, Shenyang 110004, Liaoning Province, China
Correspondence to:Yong-Lai Wang, Department of Obstetrics and Gynecology, Shengjing Hospital of China Medical University, 36 Sanhao Street, Heping District, Shenyang 110004, Liaoning Province,China. yonglai_wang@sina.com
Received:2006-12-12 Accepted:2007-01-10
Abstract
AIM:To study the reliability and tissue damage of ligasure vessel sealing system during vascular occlusion.
METHODS: Ligasure vessel sealing system was used to seal and cut off the dog blood vessel of different locations and diameters. The diameter of the vessel was measured by a caliper before and after sealing and cutting off, respectively. The bursting pressures were recorded after sealed, and the vascular pathology was performed after the operation.
RESULTS: A total of 12 vessels were sealed successfully by ligasure vessel sealing system, and the success rate was 100%. The mean diameters of the sealed artery and vein were 3.26 mm (1.75-5.22 mm) and 5.42 mm (2.40-9.08 mm), while the bursting pressures were 472 mmHg (380-560 mmHg) and 378 mmHg (250-490 mmHg), respectively. The pathological changes of the sealed vessels caused by ligasure vessel sealing system mainly presented as acute thermal and pressure-induced injury.
CONCLUSION: Ligasure vessel sealing system can effectively seal the vessels, especially the ones 3 mm or larger in diameter.
Key Words: Blood vessel; Hemostasis; Ligasure vessel sealing system
Wang YL, Chen S, Su H, Liu CG. Reliability of hemostasis with ligasure vessel sealing system in experimental dogs. Shijie Huaren Xiaohua Zazhi 2007;15(8):890-892
摘要
目的: 研究结扎速(ligasure)血管闭合系统闭合实验狗血管的可靠性及组织损伤程度.
方法: 应用结扎速血管闭合系统切断和闭合实验狗体内不同部位与直径的血管, 切断和闭合血管前用游标尺测量血管直径, 闭合后测定血管破裂时的压力, 术后血管行组织病理学检查.
结果: 结扎速血管闭合系统闭合12条动、静脉, 闭合成功率均为100%, 平均动静脉直径分别是3.26 mm(1.75-5.22 mm)和5.42 mm(2.40-9.08 mm). 成功闭合血管的平均动、静脉破裂压分别是472 mmHg(380-560 mmHg), 378 mmHg(250-490 mmHg). 术后损伤的主要病理改变为组织的急性热和压力损伤.
结论: 结扎速血管闭合系统闭合直径3 mm或更粗的血管有较高的可靠性.
关键词: 血管; 止血; 结扎速血管闭合系统
王永来, 陈双, 苏华, 刘彩刚. 结扎速血管闭合系统闭合实验狗血管的可靠性. 世界华人消化杂志 2007;15(8):890-892
0 引言
随着腹腔镜技术的飞速发展, 传统的切开及止血方法已经满足不了腹腔镜手术的要求. 目前有关各种止血设备闭合血管的可靠性及组织损伤程度的实验研究少见报道. 结扎速血管闭合系统(ligasure vessel sealing system)是一种新型止血设备, 其独特之处是以组织反应发生器作为电流和电压的能量来源. 我们通过研究结扎速血管闭合系统闭合血管的可靠性及组织损伤程度, 为进行腹腔镜手术的临床医生提供科学依据 ......
您现在查看是摘要介绍页,详见PDF附件(1kb)。