超声波测量腹壁皮下脂肪方法的探讨
作者:张新华 蔡瑞康 张宝真 李志君 冯瑞娟
单位:100036 空军总医院与第四军医大学联合培养研究生(张新华);空军总医院皮肤科(蔡瑞康)、特诊科(李志君);军事医学科学院六室(张宝真);解放军306医院内九科(冯瑞娟)
关键词:脂肪;超声波;测量
中华医学美容杂志990204 摘 要:目的 为健美类化妆品的功效进行标准化的评价,探索皮下脂肪厚度的测量方法。方法 作者利用5MHz超声波,重复测量了30例志愿者的腹部皮下脂肪厚度,并对重复性进行分析。结果 超声波测量腹部皮下脂肪厚度的信度(精确度)、效度(准确度)、相应度(变应度)均较好。结论 日内变异和日间变异均不超过10%,可在实践中应用。
Uritrosonic image measurement methodologic study of the
, http://www.100md.com
subcutanous fatty thickn ess
Zhang Xinhua,Cai Ruikang,Zhang Baozheng,et al.
Air Force General Hos pital Derpartment of Dermatology ,Beijing 100036
Abstract:Objective In order to study the measurement metho dology of the subcutanous fatty thickness.Methods We meased the abdominl subcutanous fatty thick (30 case ,4 point,every point 9 times) by 5 MH z urltrosonic wave and analysised the reappearance by paired t-test and analysi s of variance and the distribution of coefficient of variance.Results This method is better in reliablility , validity, responsibility to chan ge.Conclusion The result suggested that we can measure the subc utanous fatty thick by this method.
, 百拇医药
Key words:Fat Urltrosonic image Measurement
随着生活质量的提高,肥胖成为备受关注的课题,用于描述肥胖程度及消长的指标很多。如多个部位的皮皱厚度、腹围臀围、体重等众多指标,均有助于对肥胖的描述,也常用于表达减肥药物的效果。但这些指标均不够精确,我们应用超声波直接测量了30例志愿者的腹部皮下脂肪厚度。现介绍如下:
一般资料
1.志愿者为本院工作人员30名,男女各15名。年龄21~63岁,平均39.0±10.0岁。
2.纳入标准 ①平素体健者。②年龄18~60岁。③能够积极配合者。
3.排除标准具备下列一项者:①糖尿病、甲状腺机能异常等代谢性、内分泌疾病患者。②器质性疾病所致消耗者。③各种原因导致的水肿者。④近期用过或正在使用减肥药物、利尿药物、皮质类固醇激素等。⑤妊娠女性。
, http://www.100md.com
4.ATL超9B超仪,7CM、5MHZ探头。
测量方法
1.专人专项操作进行测量。
2.要求被测量者在测量期间生活如常,不服用利尿剂、激素等影响代谢的药物。
3.采用5MHz超声波探头,志愿者平卧,测量腹壁脐上、脐下、脐左、脐右5cm处4点皮下脂肪厚度。每日每点测3次,取均值,测量间隔时,要求被测者下床活动5分钟。
4.初次测量后第5、10日各重复测量一次。
5.测量均在下午14:00~16:00进行。
6.统计方法:配对T检验。变异度(变异系数)分布分析。
, 百拇医药
结果
30例志愿者的腹部皮下脂肪厚度情况见(表1,2)。
表1 15例男性志愿者腹壁皮下脂肪三次四点测量结果 日期
脐上
脐下
脐左
脐右
第一次
16.7±4.9
16.2±5.8
16.3±5.8
, http://www.100md.com
17.3±5.8
第二次
17.6±5.4
17.0±5.9
16.8±5.7
17.8±5.5
第三次
17.3±5.3
17.4±6.2
17.0±5.9
18.9±5.7
表2 15例女性志愿者腹壁皮下脂肪三次四点测量结果 日期
, http://www.100md.com
脐上
脐下
脐左
脐右
第一次
20.9±5.3
22.3±5.8
22.6±6.4
24.1±6.9
第二次
21.8±6.0
24.1±6.9
23.4±6.6
, http://www.100md.com
23.9±6.3
第三次
22.1±5.5
25.4±7.8
24.4±6.8
25.8±7.0
第0、5、10日测量结果(见表3,4)。表3 15例男性志愿者配对t检验结果
脐上
脐下
脐左
脐右
第一次:第二次
, 百拇医药
0.583±1.581
0.840±1.514*
0.460±0.926
0.493±1.727
第一次:第三次
0.573±1.843
1.233±1.093^
0.653±0.926*
1.607±2.359*
第三次:第二次
0.280±1.798
, http://www.100md.com
0.260±1.166
0.193±1.111
1.113±1.421^
d±S *P<0.05 ^P<0.01表4 15例女性志愿者配对t检验结果
脐上
脐下
脐左
脐右
第一次:第二次
0.820±1.760
1.840±1.934^
, 百拇医药
1.760±1.769
0.227±1.609
第一次:第三次
1.200±1.570*
3.113±2.973^
1.813±1.455^
1.653±1.611^
第三次:第二次
0.380±1.785
1.293±2.162*
1.053±1.971
, 百拇医药
1.880±2.479*
d±S *P<0.05 ^P<0.01
由表1,2可见第一次、第二次测量结果,除男女脐下测量点有显著差别外,其余各点均无显著差异。
第一次、第三次测量结果比较,除男上测量点有显著差异外,其它各点均无显著差异。
第二次、第三次测量结果比较,除女右、女下、男右测量点外,均无显著差异。
日内测量变异系数CV为0.0433±0.0316,95%分布范围0~0.12,中位数MD=0.0356,呈偏峰分布(图1)。
, 百拇医药
图1 日内变异频数分布 MD=0.0356 =0.0433±0.0316
日间测量变异系数CV为0.0603±0.0346、中位数MD=0.0557,分布近似正态。95%分布范围0.0076~0.1282(图2)。组内相关系数R均大于0.90(表5)。
图2 日间变异频数分布 MD=0.0557 =0.0603±0.0340
表5 30例男女志愿者四点组内相关系数 性别
, 百拇医药
脐上
脐下
脐左
脐右
男性
0.9408
0.9796
0.9834
0.9343
女性
0.9477
0.9077
0.9542
, http://www.100md.com
0.9409
针对任一点的变异系数(总变异)0.0693±0.0295、中位数MD=0.0663也近似正态分布。95%分布范围0.0114~0.1270(图3)。
图3 总变异频数分布 MD=0.06625 =0.069±0.0295
讨 论
腹部皮下脂肪厚度的测量方法,过去一直以测量皮皱厚度为主。这种测量方法受皮肤厚度、皮肤弹性、测量力度影响,不能直观反映皮下脂肪厚度。皮下脂肪厚度在生物个体间有明显差异,并且处于近似流体状态,使其无论在体内、体外、介入与否,均不能用直观手段测量。
, 百拇医药
应用超声波测量腹壁皮下脂肪厚度,个体测量值的变异主要来源于以下几个方面:①体位的轻微改变、呼吸运动、饮食及腹壁形状的改变。②饮食、运动等代谢形成的生理变化。③测量误差。表现在相同条件下进行重复测量结果与真值的差异,或是几个测量值的近似程度。本实验的测量误差可能来自测量点位置变化,操作者使用探头力度变化,仪器状态变化。日内测量变异偏小,X=0.0433±0.0316。日间测量变异X=0.0603±0.0346,较日内变异稍大,但分布近似正态。总变异X=0.0693±0.0295包括日内变异、日间变异的全部内涵,分布也近似正态,95%分布范围(0.0114~0.1270)同日间变异相近。配对t检验显示,虽然某些测量点日间测量值有显著差异,但变异系数为0.0693±0.0295,最大变异系数<0.13,从总体来看,测量的重复性较好。我们可以 从准确度、精确度、响应度三方面探讨 。
1.准确度(效度):①生物物理理论可以证明,5MHz超声波穿透力弱,但分辨率高,可以清晰分辨皮肤、皮下脂肪、肌肉层次,在腹壁能够反映出解剖结构,故这种测量方法的涵义效度是明确的。②因为不可能知道皮下脂肪厚度的真值,所以没有‘金标准’的标准效度。③本实验没有其它方法比较,所以结构效度不能推断。今后如进一步研究,应增加其它测量方法,相信对结构效度能有进一步了解。
, http://www.100md.com
2.精确度,即可靠性或信度,重复测量的相似程度。主要受随机因素影响,本次测量体系中,检验者为高级技师,仪器性能稳定,被测者整体集合相对不变,信度主要来自随机因素,评价观测者重复测量的一个指标是组内相关系数(R)。因为差异来源于测量者、被测者和随机误差三个方面,所以是复杂重复信度。其组内相关系数大,信度就好,反映了不同测量结构的相似程度。本实验中,组内相关系数值均在0.90以上,相对美国DrMarks年的三角肌皮皱厚度测量(R≥0.80)的信度要好。复杂重复测量的组内相关系数的总体参数ρ的(1-α)置信限,目前统计软件中,没有简单算法。假如今后在实际测量中要求必须达到R≥R*,我们可以对同一观察对象进行重复测量,用均数作为测量结果,测量次数估计值M=R*(1-R)/R(1-R*)。
本次实验没有加入干扰因素(节食、运动、减肥药物等),对这种测量方法的响应度不详,今后可以进一步研究。
收稿:1998-08-05 修回:1999-05-08, http://www.100md.com
单位:100036 空军总医院与第四军医大学联合培养研究生(张新华);空军总医院皮肤科(蔡瑞康)、特诊科(李志君);军事医学科学院六室(张宝真);解放军306医院内九科(冯瑞娟)
关键词:脂肪;超声波;测量
中华医学美容杂志990204 摘 要:目的 为健美类化妆品的功效进行标准化的评价,探索皮下脂肪厚度的测量方法。方法 作者利用5MHz超声波,重复测量了30例志愿者的腹部皮下脂肪厚度,并对重复性进行分析。结果 超声波测量腹部皮下脂肪厚度的信度(精确度)、效度(准确度)、相应度(变应度)均较好。结论 日内变异和日间变异均不超过10%,可在实践中应用。
Uritrosonic image measurement methodologic study of the
, http://www.100md.com
subcutanous fatty thickn ess
Zhang Xinhua,Cai Ruikang,Zhang Baozheng,et al.
Air Force General Hos pital Derpartment of Dermatology ,Beijing 100036
Abstract:Objective In order to study the measurement metho dology of the subcutanous fatty thickness.Methods We meased the abdominl subcutanous fatty thick (30 case ,4 point,every point 9 times) by 5 MH z urltrosonic wave and analysised the reappearance by paired t-test and analysi s of variance and the distribution of coefficient of variance.Results This method is better in reliablility , validity, responsibility to chan ge.Conclusion The result suggested that we can measure the subc utanous fatty thick by this method.
, 百拇医药
Key words:Fat Urltrosonic image Measurement
随着生活质量的提高,肥胖成为备受关注的课题,用于描述肥胖程度及消长的指标很多。如多个部位的皮皱厚度、腹围臀围、体重等众多指标,均有助于对肥胖的描述,也常用于表达减肥药物的效果。但这些指标均不够精确,我们应用超声波直接测量了30例志愿者的腹部皮下脂肪厚度。现介绍如下:
一般资料
1.志愿者为本院工作人员30名,男女各15名。年龄21~63岁,平均39.0±10.0岁。
2.纳入标准 ①平素体健者。②年龄18~60岁。③能够积极配合者。
3.排除标准具备下列一项者:①糖尿病、甲状腺机能异常等代谢性、内分泌疾病患者。②器质性疾病所致消耗者。③各种原因导致的水肿者。④近期用过或正在使用减肥药物、利尿药物、皮质类固醇激素等。⑤妊娠女性。
, http://www.100md.com
4.ATL超9B超仪,7CM、5MHZ探头。
测量方法
1.专人专项操作进行测量。
2.要求被测量者在测量期间生活如常,不服用利尿剂、激素等影响代谢的药物。
3.采用5MHz超声波探头,志愿者平卧,测量腹壁脐上、脐下、脐左、脐右5cm处4点皮下脂肪厚度。每日每点测3次,取均值,测量间隔时,要求被测者下床活动5分钟。
4.初次测量后第5、10日各重复测量一次。
5.测量均在下午14:00~16:00进行。
6.统计方法:配对T检验。变异度(变异系数)分布分析。
, 百拇医药
结果
30例志愿者的腹部皮下脂肪厚度情况见(表1,2)。
表1 15例男性志愿者腹壁皮下脂肪三次四点测量结果 日期
脐上
脐下
脐左
脐右
第一次
16.7±4.9
16.2±5.8
16.3±5.8
, http://www.100md.com
17.3±5.8
第二次
17.6±5.4
17.0±5.9
16.8±5.7
17.8±5.5
第三次
17.3±5.3
17.4±6.2
17.0±5.9
18.9±5.7
表2 15例女性志愿者腹壁皮下脂肪三次四点测量结果 日期
, http://www.100md.com
脐上
脐下
脐左
脐右
第一次
20.9±5.3
22.3±5.8
22.6±6.4
24.1±6.9
第二次
21.8±6.0
24.1±6.9
23.4±6.6
, http://www.100md.com
23.9±6.3
第三次
22.1±5.5
25.4±7.8
24.4±6.8
25.8±7.0
第0、5、10日测量结果(见表3,4)。表3 15例男性志愿者配对t检验结果
脐上
脐下
脐左
脐右
第一次:第二次
, 百拇医药
0.583±1.581
0.840±1.514*
0.460±0.926
0.493±1.727
第一次:第三次
0.573±1.843
1.233±1.093^
0.653±0.926*
1.607±2.359*
第三次:第二次
0.280±1.798
, http://www.100md.com
0.260±1.166
0.193±1.111
1.113±1.421^
d±S *P<0.05 ^P<0.01表4 15例女性志愿者配对t检验结果
脐上
脐下
脐左
脐右
第一次:第二次
0.820±1.760
1.840±1.934^
, 百拇医药
1.760±1.769
0.227±1.609
第一次:第三次
1.200±1.570*
3.113±2.973^
1.813±1.455^
1.653±1.611^
第三次:第二次
0.380±1.785
1.293±2.162*
1.053±1.971
, 百拇医药
1.880±2.479*
d±S *P<0.05 ^P<0.01
由表1,2可见第一次、第二次测量结果,除男女脐下测量点有显著差别外,其余各点均无显著差异。
第一次、第三次测量结果比较,除男上测量点有显著差异外,其它各点均无显著差异。
第二次、第三次测量结果比较,除女右、女下、男右测量点外,均无显著差异。
日内测量变异系数CV为0.0433±0.0316,95%分布范围0~0.12,中位数MD=0.0356,呈偏峰分布(图1)。
, 百拇医药
图1 日内变异频数分布 MD=0.0356 =0.0433±0.0316
日间测量变异系数CV为0.0603±0.0346、中位数MD=0.0557,分布近似正态。95%分布范围0.0076~0.1282(图2)。组内相关系数R均大于0.90(表5)。
图2 日间变异频数分布 MD=0.0557 =0.0603±0.0340
表5 30例男女志愿者四点组内相关系数 性别
, 百拇医药
脐上
脐下
脐左
脐右
男性
0.9408
0.9796
0.9834
0.9343
女性
0.9477
0.9077
0.9542
, http://www.100md.com
0.9409
针对任一点的变异系数(总变异)0.0693±0.0295、中位数MD=0.0663也近似正态分布。95%分布范围0.0114~0.1270(图3)。
图3 总变异频数分布 MD=0.06625 =0.069±0.0295
讨 论
腹部皮下脂肪厚度的测量方法,过去一直以测量皮皱厚度为主。这种测量方法受皮肤厚度、皮肤弹性、测量力度影响,不能直观反映皮下脂肪厚度。皮下脂肪厚度在生物个体间有明显差异,并且处于近似流体状态,使其无论在体内、体外、介入与否,均不能用直观手段测量。
, 百拇医药
应用超声波测量腹壁皮下脂肪厚度,个体测量值的变异主要来源于以下几个方面:①体位的轻微改变、呼吸运动、饮食及腹壁形状的改变。②饮食、运动等代谢形成的生理变化。③测量误差。表现在相同条件下进行重复测量结果与真值的差异,或是几个测量值的近似程度。本实验的测量误差可能来自测量点位置变化,操作者使用探头力度变化,仪器状态变化。日内测量变异偏小,X=0.0433±0.0316。日间测量变异X=0.0603±0.0346,较日内变异稍大,但分布近似正态。总变异X=0.0693±0.0295包括日内变异、日间变异的全部内涵,分布也近似正态,95%分布范围(0.0114~0.1270)同日间变异相近。配对t检验显示,虽然某些测量点日间测量值有显著差异,但变异系数为0.0693±0.0295,最大变异系数<0.13,从总体来看,测量的重复性较好。我们可以 从准确度、精确度、响应度三方面探讨 。
1.准确度(效度):①生物物理理论可以证明,5MHz超声波穿透力弱,但分辨率高,可以清晰分辨皮肤、皮下脂肪、肌肉层次,在腹壁能够反映出解剖结构,故这种测量方法的涵义效度是明确的。②因为不可能知道皮下脂肪厚度的真值,所以没有‘金标准’的标准效度。③本实验没有其它方法比较,所以结构效度不能推断。今后如进一步研究,应增加其它测量方法,相信对结构效度能有进一步了解。
, http://www.100md.com
2.精确度,即可靠性或信度,重复测量的相似程度。主要受随机因素影响,本次测量体系中,检验者为高级技师,仪器性能稳定,被测者整体集合相对不变,信度主要来自随机因素,评价观测者重复测量的一个指标是组内相关系数(R)。因为差异来源于测量者、被测者和随机误差三个方面,所以是复杂重复信度。其组内相关系数大,信度就好,反映了不同测量结构的相似程度。本实验中,组内相关系数值均在0.90以上,相对美国DrMarks年的三角肌皮皱厚度测量(R≥0.80)的信度要好。复杂重复测量的组内相关系数的总体参数ρ的(1-α)置信限,目前统计软件中,没有简单算法。假如今后在实际测量中要求必须达到R≥R*,我们可以对同一观察对象进行重复测量,用均数作为测量结果,测量次数估计值M=R*(1-R)/R(1-R*)。
本次实验没有加入干扰因素(节食、运动、减肥药物等),对这种测量方法的响应度不详,今后可以进一步研究。
收稿:1998-08-05 修回:1999-05-08, http://www.100md.com