中国人仿真头颈部体模的解剖参数及CT值研究
作者:高向涛 林大全 颜海婴 王江
单位:高向涛 颜海婴 (成都, 四川省肿瘤医院CT室 610041);林大全 王江 (四川联合大学人机工程中心)
关键词:仿真体模;头部;解剖参数;CT值
中华放射医学与防护杂志980617 目的 探讨中国头颈部体模的解剖结构,制作工艺和材料的合理性,以及实际应用的研究。方法 以中国“成都头颈部剂量体模”与美国RSD头部体模在解剖参数进行对比并与活体脑、软组织和头颅骨各部的CT值、质量密度、有效原子序数理论值、实测值等做比较并进行CT扫描辐射剂量分布的测定。结果 中外体模解剖参数比较差异有显著性,作为活体组织的替代材料其等效误差在5%,达到国际标准A级。结论 中国头颈部体模能够代替活体应用于放射诊断治疗和防护的研究。
Dissection parameters and CT values of a Chinese head and neck phantom.
, 百拇医药
Gao Xiangtao, Lin Daquan, Yan Haiying, et al Sichuan Cancer Hospital, Chengdu, 610041 China.
Objective To study the dissection structure of a Chinese head and neck phantom and to evaluate the reliability of technology and materials of phantom manufacture as well as its clinical appliability. Methods The phantom was compared with the head phantom of RSD,USA,and Chinese “Chengdu head dosimetric phantom” and with CT values,mass density,theoretical and experimental effective atomic numbers of in vivo human soft tissues,brain and parts of the skull.CT-scan dose distribution was also mreasured.Resultls The experimental results showed that there were remarkable differences between the dissection parameters of the Chinese phantom and the American one.The equivalent error as the substitute for human head and neck is under 5%,which reaches Class A of the international standard. Conclusions The Chinese head and neck phantom can be used as a substitute for the living body radiological diagnosis,radiological treatment and protection.
, http://www.100md.com
Key words Anthropomorphic phantom Head Dissection parameter CT values
组织等效材料(Tissue Equivalent Material)和仿真人体模型(Anthropomorphic phantom)经历了半个多世纪的发展,目前已作为“人体替身”在放射医学及其他高新技术领域成功地应用[1]。鉴于电离辐射既有诊断、治疗疾病有利的一面,又有损伤正常组织的有害影响。80年代以来,世界卫生组织(WHO),国际放射防护委员会(ICRP),国际医学物理学组织(IOMP)等,提出了医疗照射质量保证(Quality assurance)和质量控制(Quality control)概念,以期以最少的投入、最小的代价获得最佳诊断和治疗的效果。医学实践表明,仿真人体模型的应用有利于实现医学图像视觉评价与物理参数分析相结合;有利于实现体内外剂量分布的理论计算和真实模拟修正相结合;有利于实现放射设备经验管理为量化管理。本文对用于X,γ刀脑部肿瘤治疗和鼻咽癌治疗的头颈部体模与美国RSD公司生产的头部体模,从解剖参数特征、各相关组织CT值的对比分析;又对头颈部体模进行了CT断层扫描,脑部的受照剂量分布进行了初步研究。
, http://www.100md.com
1 材料方法和原理
1.1 实验用CDD-V型X,γ刀头颈部体模(由四川联合大学人机工程中心研究提供)按中国人体参数制作,用含高磷高钙的树脂制成高仿真人工颅骨,又用EVA聚乙烯-醋酸乙烯酯共聚物为基材的热塑性复合材料,模拟了脑组织左右半球、眼晶体,用高含水量稳定凝胶体制作舌、粘膜和鼻咽腔隙[2]。又用美国RSD公司生产的RSD头部模型相比较,对骨骼和肌肉组织不同的部位CT值进行比较实测,以期对组织的辐射等效误差级别与欧美人种体模的差异,及工艺的可行性临床实用性进行评价。
1.2 用以色列产EXEL 2 000 Sprint型CT机断层照像技术研究体模形态特征,CT值与活体组织等效差异。X射线CT成像技术,是通过测定不同角度检测透过的被照物体X射线强度变化,借助于计算机和数学方法,而重建照射物体内部的结构图像,这些图像的建立是基于物质对X射线线性衰减系数μ(Linear attenuation coefficient)的数字图转化为灰度级差异图像。人体是一个非均匀的介质,它由骨骼、肌肉、脂肪、脏器、腔隙构成,吸收和散射的多少与组织的元素的组成、质量密度ρ、电子密度ne和X射线的能量Exy密切相关,其射线束的衰减可表达为介质各组成元素质量衰减系数μ/ρ(mass attenuation coefficient),组成百分额Wi,质量密度ρ,和穿行厚度d的函数形式。 (1) (2)
, 百拇医药
式中I0为入射X射线强度,Ii为X射线在I点处强度,μ/ρ为线束透过每个不同质量单元的质量衰减系数,ρi为质量密度;厚度di根据线性衰减系数可换算成CT值(Hu)[3,4]。 (3)
式中K为入射X射线与射线装置投射相关的比例系数,μ组织和μ水分别为组织和水的衰减系数。
1.3 γ射线能谱仪及计算机系统
采用Canberra多道能谱仪和SPA单光子能谱仪、核素源Pu-239。γ光子能量分别为13.6,17.2,20.2keV,Am-241光子能量59.6keV测量医疗照射能谱下,质量衰减系数μ/ρ及其相关辐射参数[5]。
, http://www.100md.com
应用13~20keV低能区间,线性衰减系数μ(x)与外层电子数与介质有效原子序数相关性曲线。从而可估算介质有效原子序数。
Y=5.18+1.91X (5
Y=7.16+1.02X (11
用电子密度ne与光子射线作用总截面δe与质量衰减系数相关性估算电子密度
μ/ρ=δe×ne (6)
μ/ρ=ne×[0.33557-0.11458×Er(MeV)]
, 百拇医药
(Er为600~1 500keV) (7)
1.4 X-CT头部体模扫描剂量的测定
在头颈部体模脑部及眼晶体各个相关部分安放热释光剂量计LiF(Mg,Cu,P)管壁为聚乙烯复合等效材料(Solid Water质量密度ρ~1.0)φ2mm×2mm,进行剂量测量,经中国测试技术研究院检定,在光子能量40~662keV范围,剂量误差小于5%。进行头部7个断面CT扫描,提供人体在进行一次头部检查时的各部位的受照剂量[6,7]。
2 结果
根据公式计算出的组织替代物的CT计算值与其CT实测值比较见表1。中国人仿真头颈部体模与美国RSD及真人实测CT值比较见表2;运用X射线CT机,以中国人仿真头颈部体模测定一次头颅CT扫描(包括定位和横断扫描),各组织受照剂量情况见表3;中国人体仿真头颈部体模与中国活体头颅(中型)及美国RSD体模外型比较见表4。
, 百拇医药
表1 中国人仿真头颈部体模材料与CT值(Hu) 组织替代物
质量密度
g/cm2
线性衰
减系数
有效原
子序数
CT计
算值
CT实
测值
偏差%
, 百拇医药
肌肉TPE/SR
1.07±0.01
0.196
7.58
66
70±4.5
4.3
颅骨VF/SR
1.64±0.1
0.4048
10.7
1225
1200±130
, http://www.100md.com
2.1
下颌骨皮质VF/SR
1.74±0.1
0.4416
13.8
1433
1400±22
2.4
脑实质EVA/SR
1.04±0.04
0.1909
7.1
, http://www.100md.com 40
38±4
5.2
脂肪EVA/SR
0.92±0.01
0.1692
6.65
-80
-80±4.3
1
牙齿VP/SR
1.78±0.1
0.4452
, 百拇医药
13.8
1516
1420±9.4
6.7
表2 中国人仿真头颈部体模
与美国头部体模真人实测CT值比较 脏器、组织
中国成人
头颅活体
中国头颈
部体模
美国RSD
头部体模
, 百拇医药
肌肉
70±10
70±5
70±5
皮肤
60±10
65±5…
颅骨
1200±200
1200±100
1150±100
下颌骨皮质
1400±200
, 百拇医药
1400±100
1300±100
下颌骨髓质
180±50…
300±50
脑实质
30±5
28±4…
脂肪
-80± 10
-80± 10…
牙齿
1450±100
, 百拇医药
1420±100
180±20
实测表明,中国头颈部体模与中国成人头颅活体相比无显著性差异。中国体模在组织材料的多样性、辐射等效均匀性和实用性优于国外同类的体模(图1~2)。
表3 中国人仿真头模CT扫描时脑
组织的辐射剂量分布 解剖部位
吸收剂量
(mGy)
百分剂量分布
(%)
右颞叶皮质
0.906
, http://www.100md.com
56.3
右颞叶白质
1.005
62.4
右基底节
0.981
60.0
脑中心
0.991
61.6
左基底节
1.000
62.2
, http://www.100md.com
左颞叶皮质
1.367
85.0
左颞叶白质
1.471
91.4
枕叶皮质
0.949
58.6
枕叶白质
0.968
60.1
胼胝体膝
, http://www.100md.com
0.907
56.4
额叶白质
1.216
75.6
额叶皮质
1.609
100(几何等中心)
左眼晶体
2.310…
右眼晶体
2.262…
注:照射条件EXEL2000 Sprint X-CT机120kV 250mAs孔径240mm扫描厚度10mm扫描时间2.1s扫描层数7层
, 百拇医药
图1 安装热释剂量计的中国头颅体模CT扫描断层
图2 美国Anderson头颅体模CT扫描断层
表4 中国体模与美国体模6项基本解剖参数的差异比较 测量指标
中国活体(1)
中国体模(2)
美国体模(3)
P值
±s
, http://www.100md.com
±s
±s
1对2
1对3
2对3
头围(mm)
562.0
46.2
564.1
27.1
545.1
30.8
, 百拇医药
0.4418
8.6×10-6
2.64×10-7
前后颈(mm)
192.0
18.2
192.6
13.8
195.9
1.9
0.7288
0.0092
, http://www.100md.com
0.0122
横径(mm)
159.3
2.3
161.2
1.3
143.4
0.1
0.0786
1.23×10-22
7.16×10-26
全面凸角(度)
, 百拇医药
148.1
0.1
148.9
0.5
136.6
0.5
0.0005
1.53×10-23
5.03×10-27
软组织面型角(度)
169.7
0.7
, 百拇医药
170.3
1.3
160.0
2.6
0.1424
4.25×10-16
7.46×10-16
注:1.P<0.05,差异有显著性 P>0.05,差异无显著性 2.n1=15,n2=13,n3=13
全面凸角及软组织面型角见图3。
实测表明:中国头颈部体模取用中国成人活体50%参数制作,二者经单因子方差分析差异无显著性,P>0.05;中国头颈部体模与美国RSD头部体模相比,头围、前后径、横径差异有显著性,特别是全面凸角、软组织面型角、角度形成鼻型突出、前后径长、横径短的头型[8]。在小野等中心照射时,解剖参数和几何位置参数差异引起源皮距的变化,可能导致脑颞叶、眼晶体积分剂量增大,脑额叶、枕叶积分剂量减少。
, http://www.100md.com
图3 全面凸角及软组织面型角
3 结果和讨论
3.1 中国头颈部体模与美国RSD体模在外型上差异有明显性。表现了中国人与白种人在外型上不同特征,白种人头前后径长,横径短,全面凸角和软组织面型角明显较突;而中国体模在这几方面与中国标准人差异无显著性,说明中国头颈部体模在代表中国人种方面有特殊的意义,能够指导颌面部的外型修复测量,肿瘤病灶定位。
3.2 通过对中国头颈部体模与美国RSD头部体模,活体实测CT值的比较,中国头颈部体模组织和脏器的CT值测量结果表明肌肉、皮肤、颅骨、下颌骨皮质、脑实质、脂肪、牙龈与真人差异不大于±5%。用真人和美国体模与中国体模CT值实测对比表明,针对中国人诊断和治疗来说,我国研制的体模在组织器官的多样性,材料的辐射等效性和实用性均明显优于国外类似体模。通过组织替代物的CT计算和CT实测值的比较,进一步验证了组织替代物的线性衰减系数μ,质量密度ρ与CT值的相关性函数,也说明了设计采用准元素等效法,有效原子序数法在理论上有依据及实践上可行性。
, 百拇医药
3.3 运用中国人仿真头颈部体模测定一次头颅CT扫描,获得各组织受照剂量,是中国人仿真头颈部体模实践中的运用。通过实验,量化了一次头颅CT扫描各组织受辐射的情况,为放射工作人员,放射研究人员提供了一个直接的数据,根据我国现行放射防护标准,进行一次头颅CT扫描所受辐射剂量是很低的,这对放射科CT室工作人员在自身防护,对受检病人的保护及其安全评价和对CT检查的质量保证和质量控制有一定的意义。
本文受国家自然科学基金资助 资助项目号13945004
参考文献
1 International Commission on Radiation Units and Measurements. Phantom,electron,proton and neutron interaction data for body tissuss. ICRU Report 46,1992.
, http://www.100md.com
2 International Commission on Radiation Units and Measurements. Tissue substitutes in radiation dosimeter and measurement. ICRU Report 26,1989.
3 Daquan Lin,Changqiong Zheng,Yuanping Wan,et al. Design and synthesis of radiological equivalent tissue and bone material.Polymers and Biomaterials, 1991,(3):391-396.
4 Weifeng Kuang,Daquan Lin. Formulation of X-ray phantom material. Chin Sci Bull, 1995,40(17)1486-1490.
, http://www.100md.com
5 严美琼,赵庆昌,韩其健,等.一种确定等效原子序数的新方法.医学物理, 1985, 2(3),31-32.
6 林大全,张纪淮,王星泉,等.“中国模拟人”—非均匀辐照体模研制.医学物理, 1985, 2(3)23-26.
7 吴大可,林大全,张纪淮,等.X线诊断中非均匀组织等效体模内器官吸收剂量测定.医学物理,1985,2(3)26-30.
8 Scheideman GB. Cephalometric analysis of dentofalial anomalies. Am J Orthod, 1980,78(4):404-420.
(收稿:1998-02-24 修回:1998-05-29), http://www.100md.com
单位:高向涛 颜海婴 (成都, 四川省肿瘤医院CT室 610041);林大全 王江 (四川联合大学人机工程中心)
关键词:仿真体模;头部;解剖参数;CT值
中华放射医学与防护杂志980617 目的 探讨中国头颈部体模的解剖结构,制作工艺和材料的合理性,以及实际应用的研究。方法 以中国“成都头颈部剂量体模”与美国RSD头部体模在解剖参数进行对比并与活体脑、软组织和头颅骨各部的CT值、质量密度、有效原子序数理论值、实测值等做比较并进行CT扫描辐射剂量分布的测定。结果 中外体模解剖参数比较差异有显著性,作为活体组织的替代材料其等效误差在5%,达到国际标准A级。结论 中国头颈部体模能够代替活体应用于放射诊断治疗和防护的研究。
Dissection parameters and CT values of a Chinese head and neck phantom.
, 百拇医药
Gao Xiangtao, Lin Daquan, Yan Haiying, et al Sichuan Cancer Hospital, Chengdu, 610041 China.
Objective To study the dissection structure of a Chinese head and neck phantom and to evaluate the reliability of technology and materials of phantom manufacture as well as its clinical appliability. Methods The phantom was compared with the head phantom of RSD,USA,and Chinese “Chengdu head dosimetric phantom” and with CT values,mass density,theoretical and experimental effective atomic numbers of in vivo human soft tissues,brain and parts of the skull.CT-scan dose distribution was also mreasured.Resultls The experimental results showed that there were remarkable differences between the dissection parameters of the Chinese phantom and the American one.The equivalent error as the substitute for human head and neck is under 5%,which reaches Class A of the international standard. Conclusions The Chinese head and neck phantom can be used as a substitute for the living body radiological diagnosis,radiological treatment and protection.
, http://www.100md.com
Key words Anthropomorphic phantom Head Dissection parameter CT values
组织等效材料(Tissue Equivalent Material)和仿真人体模型(Anthropomorphic phantom)经历了半个多世纪的发展,目前已作为“人体替身”在放射医学及其他高新技术领域成功地应用[1]。鉴于电离辐射既有诊断、治疗疾病有利的一面,又有损伤正常组织的有害影响。80年代以来,世界卫生组织(WHO),国际放射防护委员会(ICRP),国际医学物理学组织(IOMP)等,提出了医疗照射质量保证(Quality assurance)和质量控制(Quality control)概念,以期以最少的投入、最小的代价获得最佳诊断和治疗的效果。医学实践表明,仿真人体模型的应用有利于实现医学图像视觉评价与物理参数分析相结合;有利于实现体内外剂量分布的理论计算和真实模拟修正相结合;有利于实现放射设备经验管理为量化管理。本文对用于X,γ刀脑部肿瘤治疗和鼻咽癌治疗的头颈部体模与美国RSD公司生产的头部体模,从解剖参数特征、各相关组织CT值的对比分析;又对头颈部体模进行了CT断层扫描,脑部的受照剂量分布进行了初步研究。
, http://www.100md.com
1 材料方法和原理
1.1 实验用CDD-V型X,γ刀头颈部体模(由四川联合大学人机工程中心研究提供)按中国人体参数制作,用含高磷高钙的树脂制成高仿真人工颅骨,又用EVA聚乙烯-醋酸乙烯酯共聚物为基材的热塑性复合材料,模拟了脑组织左右半球、眼晶体,用高含水量稳定凝胶体制作舌、粘膜和鼻咽腔隙[2]。又用美国RSD公司生产的RSD头部模型相比较,对骨骼和肌肉组织不同的部位CT值进行比较实测,以期对组织的辐射等效误差级别与欧美人种体模的差异,及工艺的可行性临床实用性进行评价。
1.2 用以色列产EXEL 2 000 Sprint型CT机断层照像技术研究体模形态特征,CT值与活体组织等效差异。X射线CT成像技术,是通过测定不同角度检测透过的被照物体X射线强度变化,借助于计算机和数学方法,而重建照射物体内部的结构图像,这些图像的建立是基于物质对X射线线性衰减系数μ(Linear attenuation coefficient)的数字图转化为灰度级差异图像。人体是一个非均匀的介质,它由骨骼、肌肉、脂肪、脏器、腔隙构成,吸收和散射的多少与组织的元素的组成、质量密度ρ、电子密度ne和X射线的能量Exy密切相关,其射线束的衰减可表达为介质各组成元素质量衰减系数μ/ρ(mass attenuation coefficient),组成百分额Wi,质量密度ρ,和穿行厚度d的函数形式。 (1) (2)
, 百拇医药
式中I0为入射X射线强度,Ii为X射线在I点处强度,μ/ρ为线束透过每个不同质量单元的质量衰减系数,ρi为质量密度;厚度di根据线性衰减系数可换算成CT值(Hu)[3,4]。 (3)
式中K为入射X射线与射线装置投射相关的比例系数,μ组织和μ水分别为组织和水的衰减系数。
1.3 γ射线能谱仪及计算机系统
采用Canberra多道能谱仪和SPA单光子能谱仪、核素源Pu-239。γ光子能量分别为13.6,17.2,20.2keV,Am-241光子能量59.6keV测量医疗照射能谱下,质量衰减系数μ/ρ及其相关辐射参数[5]。
, http://www.100md.com
应用13~20keV低能区间,线性衰减系数μ(x)与外层电子数与介质有效原子序数相关性曲线。从而可估算介质有效原子序数。
Y=5.18+1.91X (5
Y=7.16+1.02X (11
用电子密度ne与光子射线作用总截面δe与质量衰减系数相关性估算电子密度
μ/ρ=δe×ne (6)
μ/ρ=ne×[0.33557-0.11458×Er(MeV)]
, 百拇医药
(Er为600~1 500keV) (7)
1.4 X-CT头部体模扫描剂量的测定
在头颈部体模脑部及眼晶体各个相关部分安放热释光剂量计LiF(Mg,Cu,P)管壁为聚乙烯复合等效材料(Solid Water质量密度ρ~1.0)φ2mm×2mm,进行剂量测量,经中国测试技术研究院检定,在光子能量40~662keV范围,剂量误差小于5%。进行头部7个断面CT扫描,提供人体在进行一次头部检查时的各部位的受照剂量[6,7]。
2 结果
根据公式计算出的组织替代物的CT计算值与其CT实测值比较见表1。中国人仿真头颈部体模与美国RSD及真人实测CT值比较见表2;运用X射线CT机,以中国人仿真头颈部体模测定一次头颅CT扫描(包括定位和横断扫描),各组织受照剂量情况见表3;中国人体仿真头颈部体模与中国活体头颅(中型)及美国RSD体模外型比较见表4。
, 百拇医药
表1 中国人仿真头颈部体模材料与CT值(Hu) 组织替代物
质量密度
g/cm2
线性衰
减系数
有效原
子序数
CT计
算值
CT实
测值
偏差%
, 百拇医药
肌肉TPE/SR
1.07±0.01
0.196
7.58
66
70±4.5
4.3
颅骨VF/SR
1.64±0.1
0.4048
10.7
1225
1200±130
, http://www.100md.com
2.1
下颌骨皮质VF/SR
1.74±0.1
0.4416
13.8
1433
1400±22
2.4
脑实质EVA/SR
1.04±0.04
0.1909
7.1
, http://www.100md.com 40
38±4
5.2
脂肪EVA/SR
0.92±0.01
0.1692
6.65
-80
-80±4.3
1
牙齿VP/SR
1.78±0.1
0.4452
, 百拇医药
13.8
1516
1420±9.4
6.7
表2 中国人仿真头颈部体模
与美国头部体模真人实测CT值比较 脏器、组织
中国成人
头颅活体
中国头颈
部体模
美国RSD
头部体模
, 百拇医药
肌肉
70±10
70±5
70±5
皮肤
60±10
65±5…
颅骨
1200±200
1200±100
1150±100
下颌骨皮质
1400±200
, 百拇医药
1400±100
1300±100
下颌骨髓质
180±50…
300±50
脑实质
30±5
28±4…
脂肪
-80± 10
-80± 10…
牙齿
1450±100
, 百拇医药
1420±100
180±20
实测表明,中国头颈部体模与中国成人头颅活体相比无显著性差异。中国体模在组织材料的多样性、辐射等效均匀性和实用性优于国外同类的体模(图1~2)。
表3 中国人仿真头模CT扫描时脑
组织的辐射剂量分布 解剖部位
吸收剂量
(mGy)
百分剂量分布
(%)
右颞叶皮质
0.906
, http://www.100md.com
56.3
右颞叶白质
1.005
62.4
右基底节
0.981
60.0
脑中心
0.991
61.6
左基底节
1.000
62.2
, http://www.100md.com
左颞叶皮质
1.367
85.0
左颞叶白质
1.471
91.4
枕叶皮质
0.949
58.6
枕叶白质
0.968
60.1
胼胝体膝
, http://www.100md.com
0.907
56.4
额叶白质
1.216
75.6
额叶皮质
1.609
100(几何等中心)
左眼晶体
2.310…
右眼晶体
2.262…
注:照射条件EXEL2000 Sprint X-CT机120kV 250mAs孔径240mm扫描厚度10mm扫描时间2.1s扫描层数7层
, 百拇医药
图1 安装热释剂量计的中国头颅体模CT扫描断层
图2 美国Anderson头颅体模CT扫描断层
表4 中国体模与美国体模6项基本解剖参数的差异比较 测量指标
中国活体(1)
中国体模(2)
美国体模(3)
P值
±s
, http://www.100md.com
±s
±s
1对2
1对3
2对3
头围(mm)
562.0
46.2
564.1
27.1
545.1
30.8
, 百拇医药
0.4418
8.6×10-6
2.64×10-7
前后颈(mm)
192.0
18.2
192.6
13.8
195.9
1.9
0.7288
0.0092
, http://www.100md.com
0.0122
横径(mm)
159.3
2.3
161.2
1.3
143.4
0.1
0.0786
1.23×10-22
7.16×10-26
全面凸角(度)
, 百拇医药
148.1
0.1
148.9
0.5
136.6
0.5
0.0005
1.53×10-23
5.03×10-27
软组织面型角(度)
169.7
0.7
, 百拇医药
170.3
1.3
160.0
2.6
0.1424
4.25×10-16
7.46×10-16
注:1.P<0.05,差异有显著性 P>0.05,差异无显著性 2.n1=15,n2=13,n3=13
全面凸角及软组织面型角见图3。
实测表明:中国头颈部体模取用中国成人活体50%参数制作,二者经单因子方差分析差异无显著性,P>0.05;中国头颈部体模与美国RSD头部体模相比,头围、前后径、横径差异有显著性,特别是全面凸角、软组织面型角、角度形成鼻型突出、前后径长、横径短的头型[8]。在小野等中心照射时,解剖参数和几何位置参数差异引起源皮距的变化,可能导致脑颞叶、眼晶体积分剂量增大,脑额叶、枕叶积分剂量减少。
, http://www.100md.com
图3 全面凸角及软组织面型角
3 结果和讨论
3.1 中国头颈部体模与美国RSD体模在外型上差异有明显性。表现了中国人与白种人在外型上不同特征,白种人头前后径长,横径短,全面凸角和软组织面型角明显较突;而中国体模在这几方面与中国标准人差异无显著性,说明中国头颈部体模在代表中国人种方面有特殊的意义,能够指导颌面部的外型修复测量,肿瘤病灶定位。
3.2 通过对中国头颈部体模与美国RSD头部体模,活体实测CT值的比较,中国头颈部体模组织和脏器的CT值测量结果表明肌肉、皮肤、颅骨、下颌骨皮质、脑实质、脂肪、牙龈与真人差异不大于±5%。用真人和美国体模与中国体模CT值实测对比表明,针对中国人诊断和治疗来说,我国研制的体模在组织器官的多样性,材料的辐射等效性和实用性均明显优于国外类似体模。通过组织替代物的CT计算和CT实测值的比较,进一步验证了组织替代物的线性衰减系数μ,质量密度ρ与CT值的相关性函数,也说明了设计采用准元素等效法,有效原子序数法在理论上有依据及实践上可行性。
, 百拇医药
3.3 运用中国人仿真头颈部体模测定一次头颅CT扫描,获得各组织受照剂量,是中国人仿真头颈部体模实践中的运用。通过实验,量化了一次头颅CT扫描各组织受辐射的情况,为放射工作人员,放射研究人员提供了一个直接的数据,根据我国现行放射防护标准,进行一次头颅CT扫描所受辐射剂量是很低的,这对放射科CT室工作人员在自身防护,对受检病人的保护及其安全评价和对CT检查的质量保证和质量控制有一定的意义。
本文受国家自然科学基金资助 资助项目号13945004
参考文献
1 International Commission on Radiation Units and Measurements. Phantom,electron,proton and neutron interaction data for body tissuss. ICRU Report 46,1992.
, http://www.100md.com
2 International Commission on Radiation Units and Measurements. Tissue substitutes in radiation dosimeter and measurement. ICRU Report 26,1989.
3 Daquan Lin,Changqiong Zheng,Yuanping Wan,et al. Design and synthesis of radiological equivalent tissue and bone material.Polymers and Biomaterials, 1991,(3):391-396.
4 Weifeng Kuang,Daquan Lin. Formulation of X-ray phantom material. Chin Sci Bull, 1995,40(17)1486-1490.
, http://www.100md.com
5 严美琼,赵庆昌,韩其健,等.一种确定等效原子序数的新方法.医学物理, 1985, 2(3),31-32.
6 林大全,张纪淮,王星泉,等.“中国模拟人”—非均匀辐照体模研制.医学物理, 1985, 2(3)23-26.
7 吴大可,林大全,张纪淮,等.X线诊断中非均匀组织等效体模内器官吸收剂量测定.医学物理,1985,2(3)26-30.
8 Scheideman GB. Cephalometric analysis of dentofalial anomalies. Am J Orthod, 1980,78(4):404-420.
(收稿:1998-02-24 修回:1998-05-29), http://www.100md.com