当前位置: 首页 > 期刊 > 《医药产业资讯》 > 2011年第11期
编号:12077159
UF-1000i尿沉渣分析仪红细胞测定与显微镜检查结果的对比分析
http://www.100md.com 2011年4月15日 《中国医药导报》 2011年第11期
     [摘要] 目的:比较UF-1000i尿沉渣分析仪测定尿液红细胞与显微镜检查结果,对比两者符合率。方法:对624例尿标本同时采用UF-1000i尿沉渣分析仪测定红细胞和显微镜镜检两种方法检测。结果:UF-1000i检测尿中红细胞的阳性率比显微镜镜检阳性率高,假阳性率达到20.5%。结论:尿液标本采用UF-1000i检测红细胞时结果假阳性偏高,对此标本必须进行显微镜镜检复核。

    [关键词] UF-1000i尿沉渣分析仪;红细胞;显微镜镜检

    [中图分类号] R446.12[文献标识码]B [文章编号]1673-7210(2011)04(b)-166-02

    Comparative analysis of the results of UF-1000i urinalysis analyzer and microscope in determination of red blood cells

    LIU Ran, LIU Dianxiang, LI Yumei

    (Department of Laboratory, Daqing Second Hospital, Daqing 163461, China)

    [Abstract] Objective: To compare the compliance rate of UF-1000i urinalysis analyzer with microscope in determination of urine red blood cells. Methods: A total of 624 cases of urine specimens were collected and examined by UF-1000i urinalysis analyzer and microscope. The results of determination of the two methods were compared and analyzed. Results: In determination of urine red blood cells, the positive rate of UF-1000i urinalysis analyzer was higher than that of microscope. The false positive rate was 20.5%. Conclusion: The false positive rate of UF-1000i urinalysis analyzer is a little higher, in determination of the red blood cells in the urine specimens ,which must be reviewed by microscope. ......

您现在查看是摘要页,全文长 5156 字符